MPISuperRes-PnP: a super-resolution zero-shot plug-and-play reconstruction algorithm for magnetic particle imaging

Gapyak, Vladyslav und März, Thomas und Weinmann, Andreas (2026) MPISuperRes-PnP: a super-resolution zero-shot plug-and-play reconstruction algorithm for magnetic particle imaging. Physics in Medicine and Biology, 71 (19), Seite 195033. Institute of Physics (IOP) Publishing. doi: 10.1088/1361-6560/aea518. ISSN 0031-9155.

[img] PDF - Verlagsversion (veröffentlichte Fassung)
1MB

Offizielle URL: https://dx.doi.org/10.1088/1361-6560/aea518

Kurzfassung

Objective. Magnetic particle imaging (MPI) is a promising, emerging medical imaging modality. MPI is based on the non-linear response of magnetic nanoparticles to an applied magnetic field and does not expose the specimen to ionizing radiation. The measured signal is the voltage induced in receive coils by the particles' response. Reconstructing the particle concentration from the signal constitutes the imaging task. Even using state-of-the-art measurement-based reconstruction, the associated spatial grid is very coarse, hence super-resolution techniques are important. In this work, we propose an approach for super-resolution in MPI inspired by energy minimization. Approach. Different methods have been proposed for super-resolution in MPI, ranging from upscaling of the associated system matrix to interpolation of the reconstruction. Here we incorporate super-resolution into the reconstruction task via an energy minimization formulation. Following the plug-and-play (PnP) approach to energy minimization we derive a splitting scheme where the arising Gaussian denoising task is treated with a pre-trained learned Gaussian denoiser. Main results. We derive a super-resolution method for MPI based on a PnP approach using a pre-trained denoiser in zero-shot fashion. This way, we incorporate benefits of deep learning without training and avoid the need of training data. Further, we provide a quantitative and qualitative evaluation of the proposed method on simulated data with realistic noise. Hyper-parameters are selected via an extended parameter search. The found parameters are applied for reconstruction on real data. We qualitatively show the applicability of our method on real data (MPIData: EquilibriumModelWithAnisotropy and 2D-OpenMPI Data). Significance. The proposed method employs a deep-learning denoiser without training - thus it does not require presently scarcely available MPI training data. The denoiser behaves conservatively, i.e. no hallucination artifacts were observed. The super-resolution approach is generic such that it can be applied in future MPI contexts involving different regularizers or different imaging tasks.

elib-URL des Eintrags:https://elib.dlr.de/227806/
Dokumentart:Zeitschriftenbeitrag
Titel:MPISuperRes-PnP: a super-resolution zero-shot plug-and-play reconstruction algorithm for magnetic particle imaging
Autoren:
AutorenInstitution oder E-Mail-AdresseAutoren-ORCID-iDORCID Put Code
Gapyak, Vladyslavvladyslav.gapyak (at) dlr.dehttps://orcid.org/0009-0006-8822-9089230088530
März, Thomasthomas.maerz (at) h-da.dehttps://orcid.org/0009-0004-0226-1367NICHT SPEZIFIZIERT
Weinmann, Andreasandreas.weinmann (at) thws.dehttps://orcid.org/0000-0002-4969-7609NICHT SPEZIFIZIERT
Datum:9 Oktober 2026
Erschienen in:Physics in Medicine and Biology
Referierte Publikation:Ja
Open Access:Ja
Gold Open Access:Nein
In SCOPUS:Ja
In ISI Web of Science:Ja
Band:71
DOI:10.1088/1361-6560/aea518
Seitenbereich:Seite 195033
Verlag:Institute of Physics (IOP) Publishing
ISSN:0031-9155
Status:veröffentlicht
Stichwörter:magnetic particle imaging, superresolution, energy minimization inspired reconstruction, regularized reconstruction, plug-and-play, zero-shot denoiser, system matrix
HGF - Forschungsbereich:keine Zuordnung
HGF - Programm:keine Zuordnung
HGF - Programmthema:keine Zuordnung
DLR - Schwerpunkt:keine Zuordnung
DLR - Forschungsgebiet:keine Zuordnung
DLR - Teilgebiet (Projekt, Vorhaben):keine Zuordnung
Standort: Rhein-Sieg-Kreis
Institute & Einrichtungen:Institut für den Schutz terrestrischer Infrastrukturen > Resilienz- und Risikomethodik
Institut für den Schutz terrestrischer Infrastrukturen
Hinterlegt von: Gapyak, Vladyslav
Hinterlegt am:15 Okt 2026 12:22
Letzte Änderung:15 Okt 2026 12:22

Nur für Mitarbeiter des Archivs: Kontrollseite des Eintrags

Blättern
Suchen
Hilfe & Kontakt
Informationen
OpenAIRE Validator logo electronic library verwendet EPrints
Gestaltung Webseite und Datenbank: Copyright © Deutsches Zentrum für Luft- und Raumfahrt (DLR). Alle Rechte vorbehalten.